| Parameter | Value |
|---|---|
| Item | Specifications |
| Remark | Where measurement conditions are not specified, the conditions used were as follows: standard type measurement probe (HG-SS10C), ambient temperature of +20 ℃+68 ℉, and a clean atmosphere where water, oil, other liquids or dust does not come in contact with the equipment. |
| Stroke | 10.5 mm0.413 inor more(Note) : The position that represents "0" as an absolute value is a position where the spindle is pushed further down from the bottom dead point by 0.1 mm0.004 inor more. The term "stroke" indicates the total stroke length from the bottom dead point to the top dead point. |
| Weight | Net weight: 80 g approx. |
| Details | Air-driven type10 mm0.394 intype High precision |
| Product | Contact-Type Digital Displacement Sensor |
| Material | Body: Zinc, Holder: Stainless steel, Spindle: Tool steel, Probe (Note): Brass (body) / Ceramic (ball),, Air tube clamp: S60CM(Note) : The probe is also available as an option. |
| Accessory | Sensor head fastening wrench: 1 pc., Mounting nut: 1 pc., Seal cap: 1pc, Air tube clamp: 1 pc. |
| Resolution | 0.1 μm0.004 mil |
| Part Number | HG-S1110-AC |
| Product name | Contact-Type Digital Displacement Sensor HG-S |
| Usable fluid | Clean air (Dew point temperature: -10 ℃+14 ℉or less) |
| Sampling rate | 1 ms |
| Product Number | HG-S1110-AC |
| Applicable tube | Outside diameter: ø4 mmø0.157 in/ Inside diameter: ø2.5 mmø0.098 in |
| Measuring force | ・Measuring force changes with the air pressure used. Removing the seal cap enables the product to be used as the low measuring force type.・For the relationship between supplied air pressure and measuring force or between measuring force and pushed spindle position, see the figures below. For upward mount without a seal cap, subtract 0.2 N from the measuring force. For side mount, subtract 0.1 N from the measuring force. The following figures are only typical examples, and these relationships differ depending on the assembly accuracy of the product or the abrasion status of sealing materials. |
| Measuring range | 10 mm0.394 in(Note) : The position that represents "0" as an absolute value is a position where the spindle is pushed further down from the bottom dead point by 0.1 mm0.004 inor more. The term "stroke" indicates the total stroke length from the bottom dead point to the top dead point. |
| Grounding method | Capacitor grounding |
| Pollution degree | 2 |
| Hot swap function | Incorporated |
| Operating altitude | 2,000 m6561.68 ftor less(Note) : Do not use or store in an environment that has been pressurized to an air pressure higher than the atmospheric pressure at 0 m. |
| Operation indicator | Equipped (2-color LED: Orange / Green) |
| Tip deviation amount | 35 μm1.378 mil(typical value) |
| Applicable controller | HG-SC101(-P), HG-SC111(-P), HG-SC112(-P), HG-SC113(Note) : Be sure to use the sensor in combination with an HG-SC□ controller manufactured in or after February 2019. |
| Insulation resistance | 100 MΩ or more at 250 V DC |
| Regulatory compliance | CE Marking (EMC Directive, RoHS Directive), UKCA Marking (EMC Regulations, RoHS Regulations) |
| Working pressure range | 0.14 to 0.16 MPaWith no seal cap mounted : 0.035 to 0.045 MPa |
| Indication accuracy(P-P) | Full range: 1.0 μm0.039 milor lessLimited range: 0.5 μm0.02 milor less(any 60 μm2.362 mil) |
| Position detection method | Optical absolute linear encoder method |
| Capacity to resist pressure | 0.2 MPa |
| Environmental resistance : Protection | IP67 (IEC)With no seal cap mounted : -(Note) : Protective structure is not applicable when the sealing portions have deteriorated or become damaged. The protection level is zero when the seal cap is removed. |
| Environmental resistance : Ambient humidity | 35 to 85 % RH, Storage: 35 to 85 % RH |
| Environmental resistance : Shock resistance | 1,960 m/s2acceleration in X, Y, and Z directions three times each |
| Environmental resistance : Ambient temperature | -10 to +55 ℃+14 to +131 ℉(No dew condensation or icing allowed),Storage: -20 to +60 ℃-4 to +140 ℉ |
| Environmental resistance : Vibration resistance | 10 to 500 Hz frequency, 3 mm0.118 indouble amplitude (10 to 58 Hz), maximum acceleration 196 m/s2, (58 to 150 Hz) in X, Y, and Z directions for two hours each |
Technical specifications and performance data
Installation and operation guide