| Parameter | Value |
|---|---|
| 20 mA ... 4 mA | |
| Color | green (RAL 6021) |
| Depth | 114.5 mm |
| Width | 12.5 mm |
| Height | 99 mm |
| Certificate | CE-compliant |
| Designation | Loop-powered |
| Input signal | Temperature |
| Product type | Temperature transmitter |
| Screw thread | M3 |
| Sensor type: | -200 °C ... 850 °C (freely adjustable) |
| Configuration | Using MCR-PI-CONF-WIN configuration software package |
| Mounting type | DIN rail mounting |
| Identification | Class I, Div. 2, Groups A, B, C, D |
| Noise emission | EN 61326-1 (IEC 61326) and NAMUR NE 21 |
| Noise immunity | EN 61326-1 (IEC 61326) and NAMUR NE 21 |
| Switch-on delay | 4 s |
| Article revision | 02 |
| Housing material | Polyamide PA non-reinforced |
| Stripping length | 8 mm |
| Connection method | Pluggable screw connection |
| Mounting position | any |
| Dimensional drawing | |
| Degree of protection | IP20 |
| Output current limit | ≤ 23 mA |
| Supply voltage range | 12 V DC ... 35 V DC |
| Connection technology | 2-, 3-, 4-conductor |
| Current output signal | 4 mA ... 20 mA |
| Step response (10-90%) | < 2 s |
| Max. current consumption | < 3.5 mA |
| Configurable/programmable | Yes, programmable |
| Test voltage input/output | 2 kV AC (50 Hz, 60 s) |
| Conductor cross section AWG | 24 ... 24 |
| Conductor cross section rigid | 0.2 mm² ... 2.5 mm² |
| Output current with wire break | ≤ 3.6 mA or ≥ 21 mA (adjustable) |
| Ambient temperature (operation) | -40 °C ... 85 °C |
| Load/output load current output | max (Vsupply-12 V) / 0.023 A (current output) |
| Conductor cross section flexible | 0.2 mm² ... 2.5 mm² |
| Output current with short-circuit | ≤ 3.6 mA or ≥ 21 mA (adjustable) |
| Sensor types (RTD) that can be used | Pt 100 ; minimum measurement range 10 K |
| Transmission error resistance thermometer | 0.2 K |
| Output current range with overrange/underrange | ≤ 20.5 mA / ≥ 3.8 mA (linear increase/decrease) |
| Maximum power dissipation for nominal condition | 122.5 mW |
Technical specifications and performance data